DROP TEST QUATERNARY AMMONIUM COMPOUND (QAC) (1 drop = 1.25 ppm)
> & POLYQUAT (1 drop = 0.5 ppm)
COMPONENTS:
1 x 5385 Instruction
1 x 9012 Pipet, Calibrated 0.5 & 1.0 mL, plastic
w/brown cap
1 x 9198BR Sample Tube, Graduated, 25 mL, plastic
w/cap and brown dot
1 x R-0638BR-C Phenolphthalein Indicator (brown cap), 2 oz, DB
1 x R-0736BR-C Sulfuric Acid .6N (brown cap), 2 oz, DB
1 x R-0881-C Toluidine Blue O Indicator, 2 oz, DB
1 x R-0950-C Complexing Reagent, 2 oz
1 x R-0951-C QAC Titrating Solution, 2 oz, DB
Quaternary Ammonium Compound (QAC) Test
For 1 drop = 1.25 ppm QAC
NOTE: Run a blank using water containing no QAC. Record
drops of R-0951 QAC Titrating Solution used.
1. Rinse and fill 25 mL sample tube (#9198BR) to 25 mL
mark with water to be tested.
2. Using a 1.0 mL pipet (#9012), add 1.0 mL R-0950
Complexing Reagent.
NOTE: If sample water contains a hardness concentration
above 500 ppm, add 2.0 mL (2 x 1.0 mL) R-0950
Complexing Reagent.
3. Add 1 drop R-0638BR Phenolphthalein Indicator. Swirl
to mix. If colorless, proceed to step 4. If pink,
add R-0736BR Sulfuric Acid .6N dropwise, swirling
after each drop, until color changes from pink to
colorless.
4. Add 3 drops R-0881 Toluidine Blue O Indicator. Swirl
to mix. If QAC is present, sample should turn light
blue.
5. Add R-0951 QAC Titrating Solution dropwise, swirling
and counting after each drop, until color changes
from light blue to violet-pink. Always hold bottle
in vertical position.
NOTE: Further addition of R-0951 QAC Titrating Solution
should produce no color change.
6. Subtract drops of R-0951 QAC Titrating Solution
used in blank from drops used in sample (step 5).
Multiply by 1.25. Record as parts per million
(ppm) QAC as n-alkyl(50% C14, 40% C12, 10% C16)
dimethylbenzylammonium chloride.
NOTE: Equivalences for quaternary ammonium compounds
other than those listed must be determined by
titration with a known standard.
Polyquat Test
For 1 drop = 0.5 ppm
NOTE: Run a blank using water containing no polyquat.
Record drops of R-0951 QAC Titrating Solution
used.
1. Rinse and fill 25 mL sample tube (#9198BR) to 25 mL
mark with water to be tested.
2. Using a 1.0 mL pipet (#9012), add 1.0 mL R-0950
Complexing Reagent.
NOTE: If sample water contains hardness concentration
above 500 ppm, add 2.0 mL (2 x 1.0 mL) R-0950
Complexing Reagent.
3. Add 1 drop R-0638BR Phenolphthalein Indicator. Swirl
to mix. If colorless, proceed to step 4. If pink,
add R-0736BR Sulfuric Acid .6N dropwise, swirling
after each drop, until color changes from pink to
colorless.
4. Add 3 drops R-0881 Toluidine Blue O Indicator. Swirl
to mix. If polyquat is present, sample should turn
light blue.
5. Add R-0951 QAC Titrating Solution dropwise, swirling
and counting after each drop, until color changes
from light blue to violet-pink. Always hold bottle
in vertical position.
NOTE: Further addition of R-0951 QAC Titrating Solution
should produce no color change.
6. Subtract drops of R-0951 QAC Titrating Solution used
in blank from drops used in sample (step 5). Multiply
by 0.5. Record as parts per million (ppm) polyquat
as poly[(oxyethylenedimethyliminio)ethylene
(dimethyliminio)ethylenedichloride].
NOTE: Equivalences for polyquats other than those listed
must be determined by titration with a known
standard.
This product ships within approx. 5 to 7 business days from the state of Maryland. Listed below is the shipping transit time for 2024. UPS transit map - This product only ships UPS. We can not ship this product via Fedex or USPS.
- We can ship internationally.
- We can not ship this product to a PO box.
- If the product is Haz mat. It can not can not ship this via Air.
You will receive the tracking info after the order ships within 48 to 72 hours. We do not offer expedited shipping for this product.
If you can not order online, call in your order at 800-658-7716.
|
|
|
|
|
|
|