Taylor K-6560ChemWorld is a distributor of Taylor Technologies Test Kits and Reagents. Order directly online and save today.
REAGENTS:
2 x R-0833 DI Water
2 x R-0868-25C Conductivity Solution 2500 uS (1273 ppm NaCl)
COMPONENTS:
1 x 5848 Instruction
1 x 6560 Meter, Conductivity, Myron L DS Model EP-10
1. Rinse sample cup three times with sample to be measured.
NEVER FILL THE SAMPLE CUP BY DIPPING THE METER INTO WATER.
2. Refill sample cup with sample to be measured at least 1/4"
above the upper electrode.
3. Press the black button.
4. Read the scale value indicated by the needle to determine
parts per million (ppm) TDS.
5. Rinse sample cup three times with R-0833 DI Water (or
tap water) when finished with the meter.
METER CALIBRATION
NOTE: For calibration, TDS solution should be approximately
room temperature (25 degrees C/77 degrees F).
1. Remove the bottom cover of the meter using fingernails or
small screwdriver to loosen the front or rear edge.
2. Identify the Calibration Control knob (white dial with arrow)
connected to board so the knob can be found by touch while
calibrating.
3. Rinse sample cup three times with R-0968-1K TDS Solution
(equivalent to 1000 ppm 442TM).
4. Refill sample cup with standard solution to at least 1/4"
above the upper electrode.
5. While pressing the black button, adjust the Calibration
Control knob with finger until scale reads the value of the
standard solution.
6. Press the red button. Compare the scale reading with the
INTERNAL STANDARD value on the meter's bottom label. If
values are noticeably different, use the new value in place
of the old one (see Internal Standard).
7. Replace the bottom cover.
8. Rinse sample cup with R-0833 DI Water (or tap water).
INTERNAL STANDARD
An INTERNAL STANDARD value for the meter is on the meter's
bottom label. It is to be used between normal calibrations
as a field check of the meter's accuracy. It is not a
replacement for calibration with a standard TDS solution.
To verify the meter's calibration:
1. Press the red button.
2. The reading should match the INTERNAL STANDARD value on the
bottom label or the new value obtained from calibration with
a standard solution. Otherwise, the meter should be recalibrated.
CAUTION
DO NOT use samples hotter than 160 degrees F (71 degrees C).
Readings will not be accurate.
DO NOT splash solvents such as lacquer thinner, acetone,
benzene or chlorinated solvents on the plastic case.
DO NOT fix or modify the meter. This will void the warranty.
See Service for details.
DO NOT DIP THE METER INTO WATER. If water does get into the
meter, see Water Inside the Meter for instructions on drying.
MAINTENANCE
BATTERY CHECK
1. Remove the bottom cover of the meter using fingernails or
small screwdriver to loosen the front or rear edge.
2. Press the red button. Adjust the Calibration Control knob
to the maximum.
3. If the meter reads less than full scale, the battery should
be replaced. Detach the battery connector. Pull on vinyl
strap to remove the battery. Replace with a zinc carbon or
alkaline 9 volt battery. Reinsert vinyl strap and battery.
4. Replace the bottom cover.
SAMPLE CUP
Self-conditioning of the built-in electrodes occurs each time
the button is pressed with a sample in the sample cup. This
ensures consistent results each time. With some samples a small
downward swing of the needle is a result of this conditioning
action. This action is powerful and removes normal films of oil
and dirt. However, if very dirty samples (particularly scaling
types) are allowed to dry in the sample cup, a film will build up.
This film reduces accuracy. When there are visible films of oil,
dirt or scale in the sample cup or on the electrode, scrub them
lightly with a small brush and household cleanser. Rinse out
the cleanser and the meter is ready for accurate measurements.
WATER INSIDE THE METER
This Myron L meter is a rugged instrument and will withstand water
exposure around its sample cup, meter movement, and switches. However,
care should be taken to keep water from leaking in around the bottom
cover. It is not sealed (to prevent condensation from forming).
If water is relatively clean (i.e. tap water or better), and there
are only a few drops inside the meter, dry it as described below.
Large amounts of water, or corrosive or very dirty solutions will
almost certainly damage the meter. If this occurs, return meter
to Myron L Company for service.
TO DRY THE METER:
1. Shake excess water out of meter.
2. Dab the exposed surfaces dry with an absorbent cloth or tissue.
Avoid pushing any water into the Calibration Control or either
of the switches.
3. Air dry the meter in a warm area with the bottom cover off.
Allow several hours for thorough drying. If the water entered
through a leak in the case or sample cup, or if the meter shows
erratic readings or other unusual behavior, return it to Myron L
Company for service.
This product ships within approx. 5 to 7 business days from the state of Maryland. Listed below is the shipping transit time for 2024. UPS transit map - This product only ships UPS. We can not ship this product via Fedex or USPS.
- We can ship internationally.
- We can not ship this product to a PO box.
- If the product is Haz mat. It can not can not ship this via Air.
You will receive the tracking info after the order ships within 48 to 72 hours. We do not offer expedited shipping for this product.
If you can not order online, call in your order at 800-658-7716.
|
|
|
|
|
|
|