|
|
|
|
Taylor K-1692ChemWorld is a distributor of Taylor Technologies Test Kits and Reagents. Order directly online and save today.
TURBIDIMETRIC TEST POLYMER (0-10 PPM)
REAGENTS:
R-0830 Polymer Reagent 1
R-0831 Polymer Reagent 2
APPARATUS:
2 x 4030 Pipets, Calibrated 0.5 & 1.0 mL, plastic
1 x 5190 Instruction
1 x 6249 Filter Holder, 25 mm, Millipore
1 x 6257 Filter Discs, 2.5 um, 100/box
1 x 6260 Syringe, 30 mL, plastic
1 x 9198 Sample Tube, Graduated, 25 mL, plastic w/cap.
NOTE: It is important to adhere to the timed waiting periods
for accurate results.
NOTE: Positive interferences: Cationic surfactants,and Polyphosphates
at all levels.
Negative interferences: Alkalinity >300 ppm CaCO3;
Anionic surfactants at all levels; Azoles >10 ppm;
Fluoride >20 ppm; Glutaraldehyde >50 ppm;
Hardness >700 ppm; Iron >10 ppm; Molybdenum >10 ppm;
Orthophosphate >100 ppm; Phosphonate (ATMP, HEDP,
PBTC) >100 ppm; Silica >100 ppm; Sulfate >1000 ppm;
TDS >1500 ppm; Zinc >100 ppm.
1. Draw water to be tested into 30 mL syringe (#6260) to 30 mL mark.
2. Remove cap from bottom of a clean, dry sample tube (#9198).
3. Following the instructions below, filter water into sample tube.
Do not empty or disassemble syringe filtration system.
Retain for later use, if needed.
a. Open filter holder (#6249) and place 1 filter disc (#6257)
into filter holder. Screw filter holder together.
b. Screw filter holder onto syringe being careful not to push
on plunger.
c. Slowly depress plunger to filter water into sample tube to
25 mL mark.
4. Using a 1.0 mL pipet (#4030) add 2.5 mL of R-0830 Polymer
Reagent 1. Swirl to mix.
5. With a second 1.0 mL pipet (#4030) add 2.5 mL of R-0831
Polymer Reagent 2. Swirl to mix.
6. Place sample tube on center dot of turbidity standards and wait
5 minutes.
7. At exactly 5 minutes match the turbidity in the test sample with
closest dot on chart by looking down through solution. (Arrows
denote interpolated values between standards.) Record results
as parts per million (ppm) polyacrylic acid.
NOTE: Clean and thoroughly rinse all apparatus (sample tubes, syringe,
filter holder, test cell, and pipets) with distilled, deionized,
or tap water between tests. Pipets should be disassembled and all
parts thoroughly rinsed.
Syringe and filter holder should be cleaned as follows:
Unscrew filter holder from syringe. Open filter holder and
discard filter disc. Draw up deionized, distilled, or tap water
in syringe. Screw filter holder together and replace on syringe.
Depress plunger to clean and rinse. Repeat as necessary.
SYRINGE FILTRATION SYSTEM
COMPONENTS:
1 x 5860 Instruction
*1 x 6247 Syringe, 60 mL, plastic
1 x 6249 Filter Holder, 25 mm, Millipore
*1 x 6260 Syringe, 30 mL, plastic
Also required but not included: filter discs
*Kit may only contain one of these.
NOTE: Taylor offers one pore size in a paper filter disc:
2.5 um filter disc (#6257); and two pore sizes in MCE
membrane filter discs: 5 um filter disc (#6248) and
.45 um filter disc (#6261). The syringe filtration
system removes colloidal color and turbidity, which
would interfere when using the Midget(TM) comparator
system. This system is offered separately and in other
Taylor test kits.
1. Open filter holder (#6249) and place 1 filter disc (#6248,
#6257, or #6261) into filter holder. Screw filter holder
together and set aside.
2. Completely depress syringe plunger (#6247 or #6260). Draw
up water to be filtered by pulling plunger to the highest
mL mark.
3. Screw filter holder onto syringe being careful not to
push on plunger.
4. Slowly depress plunger to filter water into clean collection
vessel. Use filtered water as sample in test.
5. Unscrew filter holder from syringe. Open filter holder
and discard filter disc. Draw up deionized, distilled, or
tap water into syringe. Screw filter holder together and
replace on syringe. Depress plunger to clean and rinse.
Repeat as necessary.
COOLING SYSTEM & POLYMER TESTS
COMPONENTS (not Bill of Materials):
Alkalinity-total
1 x 9198G Sample Tube, Graduated, 25 mL, plastic w/cap
1 x R-0645 Total Alkalinity Indicator, DB
1 x R-0687G Sulfuric Acid .12N, DB
Chloride
1 x 9198O Sample Tube, Graduated, 25 mL, plastic w/cap
1 x R-0630 Chromate Indicator, DB
1 x R-0638O Phenolphthalein Indicator, DB
*1 x R-0686P Sulfuric Acid N, DB
1 x R-0706 Silver Nitrate Reagent, DB
Hardness-calcium
1 x 9198B Sample Tube, Graduated, 25 mL, plastic w/cap
1 x R-0011L Calcium Indicator Liquid, DB
1 x R-0653-2 Calcium Buffer, DB
1 x R-0683 Hardness Reagent, DB
Nitrite-sodium
1 x 9198R Sample Tube, Graduated, 25 mL, plastic w/cap
1 x R-0819 Ferroin Indicator, DB
1 x R-0820 CAN Solution, DB
Phosphonate
1 x 9198P Sample Tube, Graduated, 25 mL, plastic w/cap
1 x 9315 Test Paper, pH, 1.8-3.8
*1 x R-0686P Sulfuric Acid N, DB
1 x R-0697 Thiosulfate N/10, DB
1 x R-0802P XO Indicator Powder
1 x R-0803 Phosphonate Titrating Solution
1 x R-0805 Fluoride Masking Agent, DB
Polymer
1 x 5190 Instruction and Color Chart
1 x 9198 Sample Tube, Graduated, 25 mL, plastic w/cap
1 x R-0830 Polymer Reagent #1
1 x R-0831 Polymer Reagent #2
Misc.
1 x 5860 Instruction - Syringe Filtration System
1 x 5871 Instruction
*Kit contains one.
Alkalinity-total
1. Rinse and fill 25 mL sample tube (#9198G) to 25 mL mark
with water to be tested.
2. Add 3 drops R-0645 Total Alkalinity Indicator. Swirl to
mix. Sample should turn green.
3. Add R-0687G Sulfuric Acid .12N dropwise, swirling and
counting after each drop, until color changes from
green to red.
4. Multiply drops of R-0687G Sulfuric Acid .12N by 10. Record
as parts per million (ppm) total alkalinity as calcium
carbonate.
Chloride
NOTE: When sulfite content of sample water to be tested
exceeds 10 ppm, the sulfite should be oxidized to
prevent interference in test. A 25 mL water sample
is first adjusted to the appropriate pH, then 1 mL
(or 24 drops) of R-0649 3% Hydrogen Peroxide Solution
(sold separately) is added and thoroughly mixed.
Continue with the rest of the procedure.
1. Rinse and fill 25 mL sample tube (#9198O) to 25 mL mark
with water to be tested.
2. Add 2 drops R-0638O Phenolphthalein Indicator. Swirl to mix.
If sample turns red, add R-0686P Sulfuric Acid N dropwise,
swirling after each drop, until color changes from red to
colorless.
3. Add 5 drops R-0630 Chromate Indicator. Swirl to mix. Sample
should turn yellow.
4. Add R-0706 Silver Nitrate Reagent dropwise, swirling and
counting after each drop, until color changes from yellow to
a milky salmon (brick) red. Always hold bottle in vertical
position.
NOTE: Do not add enough R-0706 Silver Nitrate Reagent to give a
brown color. First change from yellow to a milky salmon
(brick) red is the endpoint.
5. Multiply drops of R-0706 Silver Nitrate Reagent by 10. Record as
parts per million (ppm) chloride.
Hardness-calcium
1. Rinse and fill 25 mL sample tube (#9198B) to 25 mL mark with
water to be tested.
2. Add 10 drops R-0653-2 Calcium Buffer. Swirl to mix.
3. Add 5 drops R-0011L Calcium Indicator Liquid. Swirl to mix. If
hardness is present, sample will turn red.
4. Add R-0683 Hardness Reagent dropwise, swirling and counting
after each drop, until color changes from red to blue. Always
hold bottle in vertical position.
5. Multiply drops of R-0683 Hardness Reagent by 10. Record as parts
per million (ppm) calcium hardness as calcium carbonate.
Nitrite-sodium
1. Rinse and fill 25 mL sample tube (#9198R) to 5 mL mark with
cooled (room temperature) water to be tested.
2. Add 4 drops R-0819 Ferroin Indicator. Swirl to mix. Sample
should turn red (orange).
3. Add R-0820 CAN Solution dropwise, swirling and counting
after each drop, until color changes from red (orange) to
blue. Always hold bottle in vertical position.
4. Multiply drops of R-0820 CAN Solution by 40. Record as parts
per million (ppm) sodium nitrite.
Phosphonate
NOTE: Run a blank using tap water. Normal blank requires about
2 drops of R-0803 Phosphonate Titrating Solution to reach
endpoint.
1. Rinse and fill 25 mL sample tube (#9198P) to 25 mL mark with
water to be tested.
2. Add: 1 drop R-0697 Thiosulfate N/10
10 drops R-0805 Fluoride Masking Agent
1 level dipper R-0802P XO Indicator Powder
Swirl to mix.
3. Adjust pH between 2.6 and 3.0:
Add 1 drop R-0686P Sulfuric Acid N. Swirl to mix. Dip test
paper (#9315) into sample in direction of arrow, for 3 seconds,
with all color zones immersed. Match indicator zone (the
unnumbered square located between 2.7 and 3.0 color standards)
with color scale. Read printed pH value. If necessary, continue
adding R-0686 Sulfuric Acid N dropwise, swirling and checking
pH with a new test paper after each drop, until a pH between
2.6 and 3.0 is obtained. Sample should be yellow.
4. Add R-0803 Phosphonate Titrating Solution dropwise, swirling
and counting after each drop, until color changes from yellow
to purple-pink. Always hold bottle in vertical position.
5. Subtract drops of R-0803 Phosphonate Titrating Solution in
blank from drops in sample (Step 4). Multiply by appropriate
conversion factor (see CONVERSION FACTORS). Record as parts per
million (ppm) phosphonate.
This product ships within approx. 5 to 7 business days from the state of Maryland. Listed below is the shipping transit time for 2024. UPS transit map - This product only ships UPS. We can not ship this product via Fedex or USPS.
- We can ship internationally.
- We can not ship this product to a PO box.
- If the product is Haz mat. It can not can not ship this via Air.
You will receive the tracking info after the order ships within 48 to 72 hours. We do not offer expedited shipping for this product.
If you can not order online, call in your order at 800-658-7716.
|
|
|
|
|
|
|